[ view Beijing Read Eagle Technology Co., Ltd
company information and other listings ]
Features: CE 0197, ISO 9001, ISO 13485 Ventilator WDH-1 Mode: A/C, A/C+SIGH,SIMV, SIMV+PSV,PCV, PSV,MMV,PEEP,CPAP Display: 10.4?TFT color screen Operation method: knob operation and touch key Technical parameter: Ventilation wave type: squire wave, deceleration wave Tidal volume: 50- 1500ml Breath frequency: 2- 60 BPM I :E ratio: 3:1- 1:3 Inspiration plateau: 0- 50% PEEP: 0- 2.0 KPa Trigger sensitivity: -0.1- 2.0 KPa, OFF Oxygen concentration: 21- 100% PSV pressure: 0 - 3.5 KPa MV: 1.0 - 30.0 L/min Monitor wave and parameters: Monitor wave: P-t, F-t Tidal volume Breath frequency I:E ratio PEEP Ppeak MV Oxygen concentration Pmean Alarm parameters and alarm setting Tidal volume upper limit: 503000 ml Tidal volume lower limit: 0 - 1000 ml Breath frequency upper limit: 10 - 100 BPM Breath frequency lower limit: 2 - 30 BPM Peak pressure upper limit: 1.0 - 6.5 KPa Peak pressure lower limit: 0 - 2.0 KPa PEEP upper limit: 0.5 - 3.0 KPa PEEP lower limit: 0- 2.0 KPa Oxygen concentration upper limit: 30% - 100% Oxygen concentration lower limit: 21% - 60% MV upper limit: 2.0 - 40 L/min MV lower limit: 0.5 - 20 L/min Apnea alarm and setting: 3 - 12 S Other function: Nebulizer Build-in battery
|
|